Vol 40 No 1 Original Article PDF

Measurement of Accommodative Amplitude Using a Wavefront Aberrometer

Robert Edward T. Ang, MD, Jennifer Aurea S. Sarmiento, MD, Jocelyn Therese M. Remo, MD, Gladness Henna A. Martinez, MD, Lilette Marie B. Canilao, MD

An individual less than 40 years of age can see near objects with ease because their natural lens still has strong accommodative capabilities.1-3 Beyond age 40, presbyopia and cataracts, which are both age-related causes of visual impairment, will eventually set in. Presbyopia is the decline in accommodation that leads to the reduced ability of the eye to focus on near objects. It is theorized that stiffening of the lens is the primary mechanism for presbyopia.4,5 This process usually becomes noticeable between ages 40 and 50 and increases with age, necessitating use of corrective lenses.

Cataract, which is the top cause of blindness worldwide in 2010,7 is formed by the compression and hardening of the lens with advancing age, leading to change in its refractive index, resulting in scattering of light.1 These changes cause blurred and reduced vision, faded color perception, glare, light sensitivity, and impaired night vision.1 During cataract surgery, the cataractous lens is removed and replaced with an intraocular lens which can be monofocal, multifocal, or accommodating. Monofocal and multifocal intraocular lenses (IOLs) are fixed lenses presumed to have no accommodating capabilities.8,10 On the other hand, accommodative IOLs were designed to simulate the accommodative properties of the natural crystalline lens.11,12 In a study on eyes implanted with 1CU® (Human Optics, Germany) accommodating IOL, accommodative amplitudes reached up to 2.0 diopters (D) using an autorefractor.13 However, in a study on Crystalens® (Bausch & Lomb, New York, USA) accommodating IOLs, the accommodative amplitudes were lower than 0.4 D in all subjects measured by a custom-developed aberrometer.14 These results were inconsistent as a result of the investigators using different methods to measure accommodation.

Accommodation has often been measured clinically using the push-up test, in which the subject focuses at a given line of text through his distance correction and is instructed to indicate when the object is moved close enough to blur the line.16 Investigations of changes in maximum accommodation with age date back to the late 1800s with studies performed by Donders who measured accommodation in 130 individuals between the ages of 10 and 80 years, using a variation of the subjective push-up technique.17 Duane presented similar data, showing decrease of accommodation with age.18 However, data on both studies presenting the norms of accommodative amplitude with age were based on the push-up technique which is subjective and does Ophthalmologynot unequivocally measure accommodation.

Recently, there have been studies comparing subjective (push up technique, minus lens blur, push down, defocus) and objective [Hartinger Coincidence Refractometer (HCR) (Carl Zeiss Meditec, Germany), Grand Seiko autorefractor (Grand Seiko Co., Ltd, Japan), and iTrace™ aberrometer (Tracey Technologies, USA)] methods of measurement.16,19-23 Results showed that the subjective methods overestimated the amplitude of accommodation, especially in the presbyopic population. Therefore, subjective tests performed on presbyopic patients may lead to erroneous conclusions. Due to a high disparity in the methods that are used, there is significant discrepancy in the results of many studies.

A non-contact ray tracing wavefront aberrometer, such as iTrace™, designed to objectively measure amplitude of accommodation, was shown to yield reproducible results.14 Aberrometry uses wavefront sensing, which is a technique of measuring the complete refractive status of an optical system.19 The iTrace™ aberrometer is based on the principle of ray tracing, which is a two-step, serial technique that uses forward projection. The ray tracing method uses a laser beam parallel to the line of sight through the pupil and measures the exact location where the laser beam reaches the retina by means of the retro-reflected light captured by reference lineal sensors.24 It then integrates these retinal spot positions to measure overall visual performance. It measures high and low order aberrations while the patient is looking at different targets separately. Through its open field measurement, it can determine the patient’s wavefront at two fixation points, namely distance and near fixation. This gives the subject the ability to look far away through the instrument, and the examiner to present stimuli at different distances to measure the subject’s accommodation power.24 Comparison of the two wavefronts at these two fixation points yields the accommodative amplitude. In clinical practice, it is obtained through a calibrated stick in which the observer can see the specific distance at which the stimulus is presented. The differences between the refractive maps can be analyzed both in quantitative and qualitative values. The quantitative values are shown as the myopic change given by the accommodation power.24 Figure 1 shows an example of the iTrace wavefront comparison display presenting the refractive map during near and far fixations separately. The refraction difference map shows the accommodative amplitude of the patient.

Figure 1. The iTrace™ aberrometry map showing the near (leftmost), far (middle), and the refraction difference (rightmost) maps. Refractive power for near and far fixation are shown separately in the upper right corner of the corresponding map. The accommodative amplitude is the amount of change in the spherical power when the eye changes focus from a distant to a near target

This study determined the accommodative amplitude of patients with different accommodative conditions using a wavefront device (iTrace™). Specifically, it compared the accommodative amplitude of phakic pre-presbyopic and presbyopic patients, and those implanted with monofocal and accommodating intraocular lenses (IOL).

METHODOLOGY
This is a prospective, cross-sectional study conducted from June to September 2014 in a private eye center.

Subjects

Phakic patients with no cataract, aged between 20-60 years, and pseudophakic patients above 60 years of age with monofocal or accommodating intraocular lenses bilaterally, were invited to participate in the study. They should have best-corrected distance vision of 20/30 or better. Pseudophakic patients should have had cataract surgery at least 3 months or longer prior to enrolment.

Excluded were patients with any ocular surgery (aside from cataract surgery), ocular abnormalities (diabetic retinopathy, glaucoma, corneal or macular pathology), and intraoperative complications such as posterior capsular rent and zonular rupture.

Informed consent was obtained in accordance with the Declaration of Helsinki and institutionally- approved human subjects protocol.

Procedures

Participants underwent a baseline ophthalmic examination, which included uncorrected distance and near vision, distance-corrected visual acuity, manifest refraction, and measurement of reading adds.

Accommodative amplitude was measured using an iTrace™ ray tracing wavefront aberrometer. The patient was seated with his head stabilized on the instrument chin rest and forehead strapped. Measurement light spot was turned off to eliminate target confusion. Fogging was disabled. For distance measurement, the room light was turned off to relax accommodation and to obtain as large a pupil as possible. The patient was given an unobstructed view of a distant target 20 feet away and a wavefront measurement was obtained. For near measurement, the room light was turned on and a light attached to the machine was aimed at a reading card held in place by a rod placed 40cm in front of the machine aperture. The eye not measured was occluded. The patient was asked to locate the smallest line in the chart that he could read clearly and asked to focus on that line. When the pupil became constricted to 2 to 3 mm, wavefront measurements were taken.

After 3 measurements each for distance and near, the near reading with the smallest pupil size was chosen to be analyzed against the far reading with the largest pupil size. The software (version 4.2) yielded a difference map providing the accommodative amplitude of the eye (Figure 1).

Statistical Analysis

Mean and standard deviation of accommodative amplitude were determined based on the measurements obtained from the iTrace™ ray tracing wavefront aberrometer. Paired t-test compared the mean amplitude of accommodation among the subgroups. A p value <0.05 was considered statistically significant. Accommodative amplitude correlation with age and refractive error in phakic and pseudophakic eyes were analyzed using linear regression.

RESULTS
One hundred and twenty (120) eyes of sixty seven (67) patients seen in a private eye center participated in the study. Eyes were divided into 2 groups: phakic and pseudophakic. Each group was further subdivided into 2 subgroups and compared. Phakic eyes were divided into prepresbyopic and presbyopic subgroups (Table 1). Group A (prepresbyopic) consisted of 23 eyes of 12 patients, aged 27.5 ± 4.7 years, with mean uncorrected distance visual acuity (UDVA) of 20/125 (logMar 0.8) and mean best-corrected distance visual acuity (BCDVA) of 20/20 (logMar 0). Group B (presbyopic) consisted of 24 eyes of 13 patients, aged 50 ± 6 years, with mean UDVA of 20/80 (logMar 0.6) and mean BCDVA of 20/20 (logMar 0). Differences between the mean UDVA and BCDVA of the two subgroups were not statistically significant (p=0.07). Refractive error of the 2 subgroups was also comparable (p=0.09).

Table 1. Demographic characteristics of phakic patients.

UDVA – uncorrected distance visual acuity; BCDVA – best corrected distance visual acuity; SD – standard deviation

Pseudophakic eyes were subdivided into those implanted with monofocal and accommodating IOLs (Table 2). Group C (monofocal IOL) consisted of 28 eyes of 16 patients, aged 69 ± 12 years, with mean UDVA of 20/30 (logMar 0.15) and mean BCDVA of 20/20 (logMar 0.03). Sixteen (16) eyes were previously implanted with Acrysof® (Alcon Laboratories, Inc, Texas, USA), 6 with Akreos™ (Bausch and Lomb, Inc. New York, USA), 1 with Softec (Lenstec, Florida, USA), 1 with Envista MX60 (Bausch and Lomb, Inc. New York, USA), and 4 with unspecified brand of monofocal IOLs. Group D (accommodating IOL Crystalens®) (Bausch and Lomb, Inc. New York, USA) consisted of 45 eyes of 26 patients, aged 67 ± 12 years, with mean UDVA of 20/20 (logMar 0.07) and BCDVA of 20/20 (logMar 0.03). Age, refractive error, and mean postoperative duration of the two subgroups were comparable (p=0.28, 0.22, and 0.09, respectively). Uncorrected visual acuity of those with accommodating IOL was better than those with monofocal IOL (p=0.04), while the mean best- corrected distance visual acuity was similar for the 2 subgroups (p=0.35).

Table 2. Demographic characteristics of pseudophakic patients.

UDVA – uncorrected distance visual acuity; BCDVA – best corrected distance visual acuity; SD – standard deviation

Accommodative Amplitude of Phakic Eyes

The mean accommodative amplitude of prepresbyopic eyes is 1.64D, significantly larger than the 0.99D measured in presbyopic eyes (Table 3). Linear regression showed a positive correlation (r2=0.926) between age and accommodative amplitude in phakic eyes (Figure 2). There was no observed correlation between refractive error (spherical equivalent) and accommodative amplitude in phakic eyes (Figure 3).

Table 3. Comparison of accommodative amplitude of the different subgroups using wavefront aberrometer


Figure 2. Relationship between age and accommodative amplitude in phakic eyes.


Figure 3. Relationship between spherical equivalent and accommodative amplitude in phakic eyes.


Figure 4. Relationship between age and accommodative amplitude in pseudophakic eyes.


Figure 5. Relationship between spherical equivalent and accommodative amplitude in pseudophakic eyes.

Accommodative Amplitude of Pseudophakic Eyes.

The mean accommodative amplitude of eyes implanted with accommodating IOLs was 0.94D, significantly higher than the 0.36D seen in eyes implanted with monofocal IOLs (Table 3). Accommodative amplitude of eyes with monofocal and accommodating IOLs did not show any correlation with age (Figure 4) nor refractive error measured as spherical equivalent (Figure 5).

DISCUSSION
The limitations of subjective methods, such as push-up techniques in measuring accommodative amplitude, are well-documented in several studies.16,20-25An objective measuring device is essential to provide accurate normative values. Clinical studies of intraocular lenses which promise to restore accommodation in presbyopic eyes should utilize objective methods of measurement for reliability and reproducibility of results. This study investigated the ability of a commercially available wavefront aberrometer (iTrace) to objectively measure the accommodative amplitude of phakic and pseudophakic Filipino patients.

Table 4. Comparison of different accommodative amplitude testing.

IOL – intraocular lens; HCR – Hartinger coincidence refractometer; OD – Right eye; OS – Left eye; NPA – near point of accommodation; RAF – Royal Air Force; DOF – depth of focus; AOA – amplitude of accommodation

Table 4 shows the different studies from 2003 to 2015 comparing the different testing methods for accommodative amplitude. In a study by Win-Hall et al,26 the mean accommodative amplitude of prepresbyopic eyes was 2.90 ± 0.99D using the iTrace™ aberrometer. This value was significantly higher than the measured mean accommodative amplitude of 1.37 ± 1.09D in prepresbyopic eyes in our study using the same device. Our value using the iTrace aberrometer was higher compared with the measurement of 0.79 ± 0.25D obtained using a custom-developed aberrometer,14 lower than the 5.0D with the Grand Seiko autorefractor,22,23 and lower than the 4.13 ± 0.09D obtained from the Hartinger coincidence refractometer.22 We tabulated the published reports to demonstrate the variability of outcomes using different devices because there is no concensus yet which is the better device to measure accommodative amplitude. Nevertheless, our study is relevant because we compared the prepresbyope to a presbyopic subgroup and we were able to demonstrate the expected decline of accommodative amplitude with increasing age.

Another important aspect of our study is the comparison of monofocal to accommodating intraocular lenses which has not been reported previously. In our study, eyes implanted with monofocal intraocular lenses showed mean accommodative amplitude of 0.36 diopters. Our measurement was similar to that observed in the previous study,23 which compared the accommodative amplitude of eyes with monofocal IOLs using iTrace™ aberrometer and Grand Seiko autorefractor.

Our study showed that eyes with the Crystalens® accommodating IOLs achieved a mean accommodative amplitude of 0.94 diopters. Our results were significantly higher compared to the value obtained from a recent study that measured accommodative amplitude in 22 eyes implanted with Crystalens® accommodating IOL using a custom-developed aberrometer.14 Compared to monofocal intraocular lenses, Crystalens-implanted eyes were expected, and shown by our results, to have a higher accommodative amplitude.

We analyzed the effect of refractive error on accommodation measurements. In a study by Abraham et al,29 they found that accommodative amplitude was higher in myopic eyes as compared to emmetropes and hypermetropes. In another study by McBrien et al,30 it was shown that late onset myopes have the largest amplitude of accommodation, as measured by near point rule. Our subjects were mostly myopic. However, we could not find a relationship between the refractive error and the accommodative amplitude in the phakic (r2 = 0.02), monofocal (r2=0.014), and Crystalens-implanted eyes (r2=0.02).

There is currently no established gold standard for measuring accommodative amplitude. The results in this study were compared to published results using different devices. Comparing different objective measures of accommodative amplitude in the same study population would provide a clearer picture of the amount of accommodation in different eye conditions.

In summary, the iTrace™ ray tracing wavefront aberrometer was able to objectively measure the differences and demonstrate the higher accommodative capabilities of prepresbyopes over presbyopes and accommodating over monofocal intraocular lenses. It can be a useful tool for accurate prescription of reading glasses, measuring residual accommo- dation to ascertain suitability for presbyopic treatments, and assessing the accommodative properties of certain IOLs.

REFERENCES

1. American Academy of Ophthalmology. Basic and Clinical Science Course Section 11. Lens and Cataract. California, USA. 2011-2012.
2. Glasser A, Campbell MCW. Presbyopia and the optical changes in the human crystalline lens with age. Vis Res 1998;38:209-229.
3. Mordi JA, Ciuffreda KJ. Dynamic aspects of accommodation: age and presbyopia. Vis Res 2004;44:591-601.
4. Hermans EA, Dubbelman M, van der Heijde GL, et al. Change in the accommodative force on the lens of the human eye with age. Vis Res 2008;48:119-126.
5. Glasser A, Campbell MCW. Biometric, optical, and physical changes in the isolated human crystalline lens with age in relation to presbyopia. Vis Res 1999;39:1991-2015.
6. Wubben T, Guerrero CM, Salum M, et al. Presbyopia: a pilot investigation of the barriers and benefits of near visual acuity correction among rural Filipino population. BMC Ophthalmol 2014;14:1-7.
7. WHO, World Health Statistics, 2010. [Retrieved April 21, 2014]. Available in http://www.who.int/blindness/causes/priority/en/index1.html
8. Berdeaux G, Meunier J, Arnould B, et al. Measuring benefits and patients’ satisfaction when glasses are not needed after cataract and presbyopia surgery: scoring and psychometric validation of the Freedom from Glasses Value Scale (FGVS). BMC Ophthalmol 2010;10:15.
9. Hayashi K, Hayashi H, Nakao F, et al. Aging changes in apparent accommodation in eyes with a monofocal intraocular lens. Am J Ophthalmol 2003;135:432–436.
10. Chen M. Accommodation in pseudophakic eyes. Taiwan J Ophthalmol 2012;2:117-121.
11. Pallikaris IG, Kontadakis GA, Portaliou DM. Real and pseudoaccommodation in accommodative lenses. J Ophthalmol 2011;2011:284961.
12. McLeod SD. Optical principles, biomechanics, and initial clinical performance of a dual-optic accommodating intraocular lens. Trans Am Ophthalmol Soc 2006;104:437–452.
13. Langenbucher A, Huber S, Nguyen NX, et al. Measurement of accommodation after implantation of an accommodating posterior chamber intraocular lens. J Cataract Refract Surg 2003;29:677–685.
14. Perez-Merino P, Birkenfeld J, Dorronsoro C, et al. Aberrometry in patients implanted with accommodative intraocular lenses�. Am J Ophthalmol 2014;157:1077-1089.
15. Takakura A, Iyer P, Adams JR, et al. Functional assessment of accommodating intraocular lenses versus monofocal intraocular lenses in cataract surgery: metaanalysis. J Cataract Refract Surg 2010;36:380–388.
16. Nemeth G, Tsorbatzoglou A, Vamosi P, et al. A comparison of accommodation amplitudes in pseudophakic eyes measured with three different methods. Eye 2008;22:65–69.
17. Donders FC. On the anomalies of accommodation and refraction of the eye. London: New Sydenham Society. 1864.
18. Duane A. Normal values of the accommodation at all ages. Reprinted from the J Am Med Assoc. 1912.
19. Maeda N. Clinical applications of wavefront aberrometry – a review. Clin Exp Ophthalmol 2009;37:118-129.
20. Wold JE, Hu A, Chen S, et al. Subjective and objective measurement of human accommodative amplitude. J Cataract Refract Surg 2003;�29:1878–1888�.
21. Ostrin LA, Glasser A. Accommodation measurements in a prepresbyopic and presbyopic population. J Cataract Refract Surg 2004;30:1435-1444.
22. Win-Hall DM, Ostrin LA, Kasthurirangan S, et al. Objective accommodation measurement with the Grand Seiko and Hartinger coincidence refractometer. Optom Vis Sci 2007;84:879-887.
23. Win-Hall DM, Glasser A. Objective accommodation measurements in pseudophakic subjects using an autorefractor and an aberrometer. J Cataract Refract Surg 2009;35:�282-290.
24. Gomez AC, Del Rey AV, Bautista CP, et al. Principles and clinical applications of ray-tracing aberrometry (Part 1). J Emmetropia 2012;3:96-110.
25. Wolffsohn JS, Hunt OA, Naroo S, et al. Objective accommodative amplitude and dynamics with the 1CU accommodative intraocular lens. Invest Ophthalmol Vis Sci 2006;�47:1230-1235�.
26. Win-Hall DM, Glasser A. Objective accommodation mea-surements in prepresbyopic eyes using an autorefractor and an aberrometer. J Cataract Refract Surg 2008;34:774-784.
27. Wolffsohn JS, Davies LN, Gupta N, et al. Mechanism of action of the Tetraflex accommodative intraocular lens. J Refract Surg 2010;26:858-862.
28. Win-Hall DM, Houser J, Glasser A. Static and dynamic accommodation measured using the WAM-5500 autorefractor. Optom Vis Sci 2010;87:873-82.
29. Abraham LM, Kuriakose T, Sivanandam V, et al. Amplitude of accommodation and its relation to refractive errors. Indian J Ophthalmol 2005;53:105.
30. McBrien NA, Millodot M. Amplitude of accommodation and refractive error. Invest Ophthalmol Vis Sci 1986;27:1187-
1190.